Description
Clean Air Engineering 's heated EMC inline filter and pump is intended for emissions testing and chemical analysis. It is important to maintain a constant temperature around 120°C to ensure that moisture, acid gas and other constituents of the gas stream do not condense before entering the analytical equipment. This can lead to equipment damage as well as loss of important data. The heated in-line filter and pump assembly helps ensure the accuracy of gas samples by providing a heated path for the gas prior to analysis. The filter, stainless steel needle valve and pump are insulated and heated. This unit includes a dual heating controller. All components are modular and therefore easy to maintain. The unit is also small and light, making it easy to transport. For use in: process monitoring, stack testing, EMC, gas and diesel sampling.
The heated filter holder in line with a heated ADI sample pump for gas sample delivery. This combination can be used with an extractive FTIR system or other hot CEMS to provide a particle-free hot sample flow for your measurement system.
Heated in-line filtration pump specifications:
Weight: 18Kg
Dimensions: 31 x 43 x 54 (L x W x H) cm
Maximum pump flow: 27 l / min
Pump amperage: 3.6-4.0 amps
Pump heater amperage: 1.25 amps
Pump power: 120/240 volts, 60/50 Hz
Pump heating power: 2 @ 75 watts with type K thermocouple
Pump: Single-head 316SS with Teflon ® diaphragm
Maximum pump heating temperature: 205°C
Filter heater power: 338.82 watts
Heater amperage: 2.82 amps
Maximum filter heater temperature: 205°C
Voltage requirements: 120 VAC
Model Filter size: 2.54 cm x 17.78 cm
Inlet and outlet connections: 3/8" stainless steel compression fitting