Description
TheAcoem Serinus 44gas analyzer measures nitrogen oxides (NO, NO2, NOx and NH3) in ambient air using its proven chemiluminescence technology. The Serinus 44 is a powerful instrument offering accurate and reliable performance, enabling governments and organizations to reduce their environmental footprint, while complying with global regulations.
The Serinus 44 uses gas-phase chemiluminescence detection for continuous analysis of nitric oxide (NO), total oxides of nitrogen (NOX) and nitrogen dioxide (NO2), as well as an external thermal catalytic converter for analysis of ammonia (NH3). The instrument consists of a pneumatic system, an NO2 to NO converter, a reaction cell, a measuring cell (PMT), an ozone generator and a PCA master controller.
Key points:
- Real-time monitoring and remote viewing of over 200 operational parameters.
- Bluetooth connectivity combined with the Android app "Serinus® Remote" to simplify programming, viewing, downloading and emailing data.
- Removable flash memory for 10 years of data storage, including up to 12 individual settings and event logs.
- Advanced graphical interface and large alphanumeric keyboard with tactile keys.
Certifications :
- LSCQA approval (France)
- U.S. Environmental Protection Agency Approval (RFNA-0809-186)
- EN TÜV approval (936/21221977/A)
- EN MCERTS approval (MC100167/06)
- Australian Standard (AS 3580.5.1-2011)
- Russian approval (56263-14)
- Approval 12/3/B/24/295-17 (Ukraine)
- Model approval JJF1361-2012 (China)
- NIER Approval AACMS-2017-01 (South Korea)
Also available:
- Ozone (O3) in ambient air : Acoem Serinus 10
- Sulphur dioxide (SO2) in ambient air : Acoem Serinus 50
- SO2 / H2S : Acoem Serinus 51
- Nitrogen dioxide (NO2) in ambient air : Acoem Serinus 60
Available for rent :
CleanAir Europe rents out the Serinus 44 analyzer for low-level measurements of NO, NO2, NOx and NH3
Features
Measuring range:
- 0 - 20 ppm (automatic changeover)
- (analog ranges 0 - 50 ppb to 0 - 20 ppm)
- (NH3 range dictated by external converter; typically ~2 ppm)
Concentration display :
- User-selectable mg/m³, μg/m³, ppm, ppb or ppt
Noise :
Lower detection limit :
- < 0,4 ppb ou 0,5 % de la lecture, la valeur la plus élevée étant retenue
Linearity :
- < 1 % de la pleine échelle
Note:
- 0.4 ppb or 0.5% of reading, whichever is greater
Zero drift :
- 24 heures : < 0,4 ppb
- 7 jours : < 0,4 ppb < 0,4 ppb
Measurement range drift :
- 7 jours : < 1 % de la valeur relevée ou 0,4 ppb, selon la valeur la plus élevée < 1 % de la valeur relevée ou 0,4 ppb, la valeur la plus élevée étant retenue
STP reference :
- 0 °C, 20 °C, 25 °C at 101.3 kPa
Sample flow :
- 900 cc/min (total for 3 channels)
Temperature range :
Power :
- 100 - 240 VAC, 50 - 60 Hz (self-tuning)
Power consumption :
- 265 VA (max. at startup)
- 190 VA (after warm-up)
Dimensions :
Bay spacing :
Weight :
Data communication/recording
- USB port (digital communication)
- Bluetooth (digital communication via Android App)
- Ethernet TCP/IP network connection (optional)
- RS232 port 1: Digital communication
- RS232 port 2: Multidrop port
- Modbus RTU/TCP protocol, advanced
- 25-pin I/O port (analog outputs, analog inputs,
analog inputs, digital status inputs, digital status outputs).
Data logging:
Removable USB flash memory stores data in the following formats
following formats:
- Instantaneous data from : 1, 3, 5, 10, 30 or 60 minutes
- Average data from : 1, 3, 5, 10, 15, 30 minutes,
1, 4, 8, 12 or 24 hours
- 10-year data storage of 12 parameters,
1 minute of data on 8 GB memory.
Accessories
- HTO-1000N NH3 converter
- External AC powered pump
- TCP/IP network port
- Bluetooth
- Rack slide kit
- Dual filter (particulate)
- High pressure zero/span valves
- Printed manual - electronic manual supplied
with analyser.