CleanAir Europe
  • Home
  • Our Services
      •   Rental
      • Maintenance
      • Calibration
  • Instruments
      • New measuring instruments
      • Gas Analysers
      • Mercury Analysers / Samplers
      • Gas Detectors
      • Particle Counters / Analysers
      • Dust Samplers
      • Gas Sampling Systems
      • Isokinetic Sampling Systems
      • Multiparameter Equipments
      • Thermal Cameras
      • Refurbished Equipments
  • Blog
  • Contact
  • About
      • About CleanAir
      • Career
  • My Quote
  • Connection
  • Search
  • Menu Menu
  • 0Basket
You are here: Home Home1 / Gas analyzers: Air quality2 / Acoem Serinus 44 ...

Acoem Serinus 44
NO / NO2 / NOx / NH3 Gas Analyzer

Request a quote
This product is already in the request for quote list.
Browse the list
Categories : Gas analyzers: Air quality, Location: Ambient air, NO / NO2 / NOX, News Tags : Gas analyzer, Chemiluminescence (CLD), NH3, NO, NO2, NOx, Air quality
  • Description
  • Features
  • Data communication/recording
  • Accessories
  • Documentation

Description

L'gas analyzer Acoem Serinus 44 gas analyzer measures nitrogen oxides (NO, NO2, NOx and NH3) in ambient air using its proven chemiluminescence technology. The Serinus 44 is a powerful instrument offering accurate and reliable performance, enabling governments and organizations to reduce their environmental footprint, while complying with global regulations.

The Serinus 44 uses gas-phase chemiluminescence detection for continuous analysis of nitric oxide (NO), total oxides of nitrogen (NOX) and nitrogen dioxide (NO2), as well as an external thermal catalytic converter for analysis of ammonia (NH3). The instrument consists of a pneumatic system, an NO2 to NO converter, a reaction cell, a measuring cell (PMT), an ozone generator and a PCA master controller.

Key points:

  • Real-time monitoring and remote viewing of over 200 operational parameters.
  • Bluetooth connectivity combined with the Android app "Serinus® Remote" to simplify programming, viewing, downloading and emailing data.
  • Removable flash memory for 10 years of data storage, including up to 12 individual settings and event logs.
  • Advanced graphical interface and large alphanumeric keyboard with tactile keys.

Certifications :

  • LSCQA approval (France)
  • U.S. Environmental Protection Agency Approval (RFNA-0809-186)
  • EN TÜV approval (936/21221977/A)
  • EN MCERTS approval (MC100167/06)
  • Australian Standard (AS 3580.5.1-2011)
  • Russian approval (56263-14)
  • Approval 12/3/B/24/295-17 (Ukraine)
  • Model approval JJF1361-2012 (China)
  • NIER Approval AACMS-2017-01 (South Korea)

Also available:

- Ozone (O3) in ambient air : Acoem Serinus 10
- Sulphur dioxide (SO2) in ambient air : Acoem Serinus 50
- SO2 / H2S : Acoem Serinus 51
- Nitrogen dioxide (NO2) in ambient air : Acoem Serinus 60

 

Available for rent :

CleanAir Europe rents out the Serinus 44 analyzer for low-level measurements of NO, NO2, NOx and NH3

Features

Measuring range:

  • 0 - 20 ppm (automatic changeover)
  • (analog ranges 0 - 50 ppb to 0 - 20 ppm)
  • (NH3 range dictated by external converter; typically ~2 ppm)

Concentration display :

  • User-selectable mg/m³, μg/m³, ppm, ppb or ppt

Noise :

  • < 0,2 ppb

Lower detection limit :

  • < 0,4 ppb ou 0,5 % de la lecture, la valeur la plus élevée étant retenue

Linearity :

  • < 1 % de la pleine échelle

Note:

  • 0.4 ppb or 0.5% of reading, whichever is greater

Zero drift :

  • 24 heures : < 0,4 ppb
  • 7 jours : < 0,4 ppb < 0,4 ppb

Measurement range drift :

  • 7 jours : < 1 % de la valeur relevée ou 0,4 ppb, selon la valeur la plus élevée < 1 % de la valeur relevée ou 0,4 ppb, la valeur la plus élevée étant retenue

STP reference :

  • 0 °C, 20 °C, 25 °C at 101.3 kPa

Sample flow :

  • 900 cc/min (total for 3 channels)

Temperature range :

  • 0 - 40 °C

Power :

  • 100 - 240 VAC, 50 - 60 Hz (self-tuning)

Power consumption :

  • 265 VA (max. at startup)
  • 190 VA (after warm-up)

Dimensions :

  • 429 x 175 x 638 mm

Bay spacing :

  • 3.5 RU

Weight :

  • 21.85 kg

Data communication/recording

- USB port (digital communication)
- Bluetooth (digital communication via Android App)
- Ethernet TCP/IP network connection (optional)
- RS232 port 1: Digital communication
- RS232 port 2: Multidrop port
- Modbus RTU/TCP protocol, advanced
- 25-pin I/O port (analog outputs, analog inputs,
analog inputs, digital status inputs, digital status outputs).

Data logging:

Removable USB flash memory stores data in the following formats
following formats:
- Instantaneous data from : 1, 3, 5, 10, 30 or 60 minutes
- Average data from : 1, 3, 5, 10, 15, 30 minutes,
1, 4, 8, 12 or 24 hours
- 10-year data storage of 12 parameters,
1 minute of data on 8 GB memory.

Accessories

- HTO-1000N NH3 converter
- External AC powered pump
- TCP/IP network port
- Bluetooth
- Rack slide kit
- Dual filter (particulate)
- High pressure zero/span valves
- Printed manual - electronic manual supplied
with analyser.

Documentation

EN Data Sheet

Similar products

  • Kunak AIR Pro
    Air Quality Monitoring Station

    Read more View details
  • ETG 6900 TDL NH3
    Ammonia Analyzer NH3 H2O

    Read more View details
  • Acoem Serinus 40
    NO / NO2 / NOX Gas Analyzer

    Read more View details
  • Thermo Scientific Model 42i
    NO / NO2 / NOX Gas Analyzer

    Read more View details

360 avenue de la Rasclave
Z.I. Braye de Cau
13821 La Penne-sur-Huveaune
France

+33(0)4 91 87 82 10
cleanair.europe@cleanair.com

CFM and Measurement Network logo

OUR EQUIPMENT

  • New measuring instruments
  • Gas Analysers
  • Mercury analyzers & samplers
  • Gas Detectors
  • Particle Counters & Analyzers
  • Dust Samplers
  • Gas Sampling Systems
  • Isokinetic Sampling Systems

    ABOUT

  • Blog
  • Career
  • Contact

OUR SERVICES

  •   Rental
  • Sale
  • Maintenance
  • Calibration
logo CB visa mastercard
Copyright - CleanAir Europe - Enfold Theme by Kriesi
  • LinkedIn
  • Facebook
  • General Conditions of Direct Sales
  • Privacy Policy
ASCO Instruments VAR18 ...Defiant Technologies VOCAM...
Scroll up
We use cookies to improve our content and services, but also to provide you with a safer experience. You can accept all cookies or manage your preferences via the "Personalise" menu. Learn more about our use of cookies and data.
CustomizeACCEPT
Manage consent

Setting cookies

Cookies classified as necessary are stored on your browser because they are essential for the basic functionality of our website. We also use third-party cookies that allow us to analyse the performance of our website. You have the option to disable these cookies. Please note, however, that disabling some of these cookies may affect your experience of browsing the site and the online services we can offer you.
Necessary cookies
These cookies are necessary for the proper functioning of our website and its services. These cookies do not store any personally identifiable information.
Performance cookies
These cookies allow us to understand and analyze key analytical performance indicators, which helps to provide a better user experience.
Advertising cookies
These cookies are used to provide visitors with relevant advertising and marketing campaigns. These cookies track visitors to the websites and collect information to provide personalized advertisements.
Save & Apply