Description
The JUM 3-300A is a proven FID as a space-saving version of the JUM VE7. It is a highly reliable 19" rack-mounted heated total hydrocarbon (VOC) analyzer. Designed for very low drift, high accuracy, sensitivity and stability. The JUM 3-300A uses a heated oven FID to avoid high hydrocarbon molecular weight loss and to provide reliable performance in analyzing high concentrations down to very low levels of trace VOCs in emissions, air and other gases.
The JUM 3-300A is a standard emission analyzer and therefore optimized in accordance with the European specifications EN-12619: 2013. Several different optimizations for non-EN- 12619: 20136 applications are available.
Confirmed by TÜV-Nord to comply with QAL1 EN 14181 and EN ISO 14956 (EU). Fully compliant with EN 12619: 2013 (EU) and EPA Method 25A and Method 503 (USA)