VA 3000 item 154

This analyzer measures the N2O levels in emissions.
It uses the principle of infra-red correlation (NDIR) to measure nitrous oxide (N2O) in emissions and in specific processes. Fitted with a CO2 filter to eliminate interference caused by this gas.
It is used after an NO2 converter gas treatment system with a yield of greater than 97%.

Specification

– Measuring range : programmable from 0 – 500 ppm to 0 – 5000 ppm
– Units : ppm, mg/m3
– Detection Limit : 1 ppm
– Internal Pump : yes
– Data storage : none (a data recorder can be added).
– Analog output : yes (0 – 1V or 4 – 20 mA)
– Communication interface : yes
– Packaged weight : 18 kg
– Number of packages : 1
– Recommended for use with : 846, 832, 820, 622, 917.

Key Points

– Reduced heat-up time
– No CO2 interference (optical CO2 filter), reduced CO interference
– Not sensitive to vibration
– Menus for setting parameters and for optical and electronic diagnostics
– Automatic range switchover