This machine will continuously measure total hydrocarbons.
The Jum 3-500 uses a flame ionization detector (FID).
It connects directly to a heated output line. It is linear up to 20,000 ppm.

Specification

– Measurement Range : 0 – 10, 10 – 100, 0 – 1000 and 0 – 20,000 ppm (linear) and above that logarithmically.
– Units : ppm
– Detection Limit : 1/1000th of the range.
– Fuel Gas: H2/He, 40/60% vol.
– FID Temperature : 190 °C
– Data storage : none (a data recorder can be added).
– Internal Pump : yes
– Analog output: 0 – 10V
– Packaged weight : 25.5 kg
– Number of packages : 1
– Recommended for use with : 622, 831, 846, 925, 925M, 928 or 928H

Key Points

– 19’’ rack, 3U.
– FID burner can be used in ambient air with an activated charcoal filter.

109L item 194

This device can simultaneously and continuously measure total hydrocarbons, methane containing hydrocarbons and non-methane hydrocarbons.
The Jum 109L uses a flame ionization detector (FID).
Total hydrocarbons are measured directly, methane hydrocarbons are measured after eliminating non-methane compounds and non-methane hydrocarbons are calculated from the difference. It connects directly to a heated output line. It is linear up to 20,000 ppm.

 

Specification

–  Measurement Range : 0 – 10, 10 – 100, 0 – 1000 and 0 – 20,000 ppm (linear) and above that logarithmically.
– Units : ppm
– Detection Limit : 1/1000th of the range.
– Fuel Gas: H2/He, 40/60% vol.
– FID Temperature : 190 °C
– Data storage: none (a data recorder can be added).
– Internal Pump : yes
– Analog output: 0 – 10V for each channel.
– Packaged weight: 22 kg
– Number of packages: 1
– Recommended for use with: 622, 831, 846, 925, 925M, 928H

Key Points

-19’’ rack, 3U.