Return to the previous page

ETG MCA 100 BIO-P ITEM 546 AMBIENT AIR AND PROCESS

This system is suitable for the measurement of the biogas in a single point, and performs continuous measurement of CH4, CO2,CO, O2 and H2S.

The MCA 100 P is the portable version of the MCA 100 Bio and has been designed for those who need to monitor the composition of the biogas at several points.

The analyzer is mounted inside a hard case impact resistant, equipped with touch screen monitor ( 5.7”), capable of recording up to 1000 data analysis, has an autonomy of 8 hours.

 

CH4, CO and CO2 board NDIR based

 

O2, H2S board electro chemical cell based

 

Specifications

– Ranges :

  • CH4 0…100 vol% (NDIR)
  • CO2 0…50% (NDIR)
  • CO 0…1% (NDIR)
  • O2 0…25 vol% (ECD)
  • H2S 0…5000 ppm (ECD)
  • Calorific Value

– Unit : %, PPM

– Data logging : Yes

– Software : Yes

– Internal pump : Yes

– Analog output : No

– Weigth : 8 kg

– Number of box : 1

– Recommended for use with: 822

Key Points :

– Real-Time analysis

– Remotable by Ethernet

– Trend of the last 30 minutes for each measured compound

– Dust filter on sample line

– Sampling pump

– Data Download .csv format by USB port

 

 

 

Servomex 1400 item 113

This analyzer measures the CO2 levels in emissions.
The Servomex 1400 uses the principle of infrared absorption.
It is used after a gas preparation system.
Set-points can be pre-defined and an alarm can be sounded if they are exceeded. A flow-rate alarm is also available.
The 0 – 10 V analog output can be used to record the data on a plotter.

Specification

– Measurement Range : 0 – 20%, 0 – 25%
– Unit : %
– Detection Limit : 0.1%
– Analog output : 0 – 10V
– Data storage : None (a data recorder can be added).
– Internal pump : No (an external pump can be added).
– Packaged weight : 15 kg
– Number of package : 1
– Recommended for use with : 846, 832, 820, 812, 622, 924

Key Points

– 19’’ rack, 4U
– 2 alarms at 2 thresholds

Gas 3000 item 547

This portable analyzer can measure CH4, CO2, O2 and H2S for biogas applications.

The GAS 3000 uses infra-red detection for CH4 and CO2 and electrochemical cells for O2 and H2S.

It is used to measure gas in the soil, to monitor landfills and waste processing sites and for biogas quality control.

It is supplied with a probe and a hydrophobic filter.

Specification

– Ranges : 0 – 100% CH4, 0 – 50% CO2, 0 – 25% O2, 0 – 5000 ppm H2S
– Units : %, ppm
– Internal Pump : yes
– Data Storage : yes
– Communication interface : yes
– Software : yes
– Recommended for use with : 622

Key Points

– Battery Life : 4 to 6 hours

Ultramat 23 item 120 121 122

This continuous gas analyzer can measure the CO and CO2 in emissions simultaneously.
The Ultramat 23 uses the principle of infrared absorption.
Several versions with different ranges are available.
The machine can be delivered with its calibration certificate on request (depending on the range).

Specification

– Measurement Ranges:
Item 120 :  0 – 0.05% and 0 – 2.5% CO and 0 – 5% and 0 – 25% CO2
Item 121 : 0 to 1000 ppm CO and 0 to 10,000 ppm CO2 (full scale)
Item 122 : 0 – 10% and 0 – 50% CO and 0 – 1% CO2
– Units : %, mg/m3
– Detection limit : 50 ppm CO / 500 ppm CO2
Item 120 : 50 ppm CO / 500 ppm CO2
Item 121 : 1 ppm CO / 1 ppm CO2
Item 122 : 0.05% ppm CO / 500 ppm CO2
– Data storage : none (a data recorder can be added).
– Internal Pump : yes
– Analog Output : 0 – 10 V
– Packaged weight : 15 kg
– Number of packages : 1
– Recommended for use with : 846, 832, 820, 622, 923, 924.

Key Points

– 19’’ rack, 4U

Q-Track item 452

This portable device measures carbon monoxide in ambient air with an electrochemical cell and measures carbon dioxide using non-dispersive infra-red detection.

It can also indicate temperature and relative humidity.

It operates on battery and mains power.

Specification

– Measurement Range : 500 ppm CO, 5000 ppm CO2
– Units : ppm, °C, %
– Temperature : from 0°C to +50°C
– Relative Humidity : 0 – 95%
– Detection Limit : 1 ppm CO, 1 ppm CO2
– Data Storage : 15000 measurement points
– Communication interface : yes
– Software : yes
– Packaged weight : 6 kg
– Number of packages: 1

 

Points clés

– Weight for user : 1 kg
– Powered by 4 AA batteries (1 set in the device, 1 set of new batteries supplied)

Gas 3100 Item 160

This analyzer measures H2.
The GEIT GAS 3100l uses the principle of the thermal conductivity of a gas.
It can also quantify CO2 to compensate for this interfering gas.

Specification

– Measurement Ranges : 0 – 20 – 100% H2, 0 – 5 – 100% CO2
– Units : %
– Detection limit : 0.01% for H2 and CO2
– Internal Pump : yes
– Analog output : yes
– Communication : yes
– Software : Yes
– Packaged weight : 20 kg
– Number of packages : 1
– Recommended for use with : 846, 832, 820, 622, 924