Return to the previous page

Servomex 1400 item 113

This analyzer measures the CO2 levels in emissions.
The Servomex 1400 uses the principle of infrared absorption.
It is used after a gas preparation system.
Set-points can be pre-defined and an alarm can be sounded if they are exceeded. A flow-rate alarm is also available.
The 0 – 10 V analog output can be used to record the data on a plotter.

Specification

– Measurement Range : 0 – 20%, 0 – 25%
– Unit : %
– Detection Limit : 0.1%
– Analog output : 0 – 10V
– Data storage : None (a data recorder can be added).
– Internal pump : No (an external pump can be added).
– Packaged weight : 15 kg
– Number of package : 1
– Recommended for use with : 846, 832, 820, 812, 622, 924

Key Points

– 19’’ rack, 4U
– 2 alarms at 2 thresholds

Servomex 1400 item 174

This machine is used to measure O2.
The Servomex 1400 uses magnetic susceptibility to continuously measure O2 levels.
The gas sample must be dry, clean and at room temperature.
It is used after a gas preparation system.
Set-points can be pre-defined and an alarm can be sounded if they are exceeded. A flow-rate alarm is also available.
The 0 – 10 V analog output can be used to record the data on a plotter.

Specification

– Measuring range : 0 – 25% and 0 – 100%.
– Units : %
– Detection Limit : 0.1%
– Data storage : none (a data recorder can be added).
– Internal pump : no (an external pump can be added).
– Analog output : 0 – 1V
– Packaged weight : 15 kg
– Number of packages : 1
– Recommended for use with : 812, 820, 831, 846, 622, 924

Key Points

– 2 alarm thresholds