Express Dilution System Item 842

Sonic nozzle dilution probe.
Ideal for reducing concentrations or drying a condensable gas by cooling.
 

Specification

– Dilution ratios 1/16 to 1/300
– Requires on site air instrument zero at 5 bar
– Requires high concentration calibration gas to set the dilution ratio
– Consumption : 6 l/min of dilutant air
– Consumption : 20 to 500 cc of calibration gas
– Packaged weight : 30 kg, 1 crate

 

Key Points

– An instrument air system is recommended.

Dusts item 311 et 311A

This isokinetic manual sampling system will continuously sample gas or vapor in a chimney while maintaining the same speed characteristics as at the sampling point. Particles are removed onto a filter. Hot gases are conditioned in a series of cold bubblers.

Condensate is recovered and a volume flowmeter (DGM) measures the dry gas flowrate.

This system can either be supplied with a heating cabinet or with a Cateco oven box which reduces its overall size.

Specification

– Sampling flow rate : depending on the isokinetic parameters < 40 liters/min
– Length of probe : 0.6 m, 0.9 m, 1.2 m, 1.5 m, 1.8 m, stainless steel (specify length with order).
– Umbilical : 7.5 m, 15 m or 30 m (specify with order).
– Set of Nozzles : 3, 4, 5, 7, 9 and 12 mm (stainless steel).
– Gas Temperature : 200 °C max (optionally 600 °C max).
– Oven Temperature : 120 °C
– Calculation Sheet : in Excel® (supplied)
– Packaged weight : 120 kg
– Number of packages : 5

 

Key Points

– Item 311A (Cateco oven box) is particularly suited for difficult sampling operations with its small probe and small oven compartment.
– This device must only be used by an operator trained in a specific methodology. The flow speed at the sampling point must be known and communicated to CleanAir Europe for assessing the feasibility of the project.

Dioxins and furans item 311D

This system is a version of the base model (dust) and has special glassware for analyzing dioxins and furans.

It can be modified for polycyclic aromatic hydrocarbons.

It can be delivered with 2 sets of glassware (item 311VD), to be ordered separately.

Specification

– Sampling flow rate : depending on the isokinetic parameters < 40 liters/min
– Length of probe : 0.6 m, 0.9 m, 1.2 m, 1.5 m, 1.8 m, stainless steel (specify length with order).
– Umbilical : 7.5m, 15m or 30 m (specify with order).
– Set of Nozzles : 3, 4, 5, 7, 9 and 12 mm (stainless steel).
– Gas Temperature : 200 °C max (optionally 600 °C max).
– Oven Temperature : 120 °C
– Calculation Sheet : in Excel® (supplied)
– Packaged weight : 120 kg
– Number of packages : 5

 

Key Points

– This device must only be used by an operator trained in a specific methodology. The flow speed at the sampling point must be known and communicated to CleanAir Europe for assessing the feasibility of the project.
 

HF & HCI item 311H

The machine can sample HF and HCl.

It is a variant of the base model (dust). It uses nozzles and a Pyrex duct with a compartment containing 4 impingers.

Specification

– Sampling flow rate : depending on the isokinetic parameters < 40 liters/min
– Length of probe : 0.6 m, 0.9 m, 1.2 m, 1.5 m, 1.8 m, stainless steel (specify length with order).
– Umbilical : 7.5m, 15m or 30 m (specify with order).
– Set of Nozzles : 3, 4, 5, 7, 9 and 12 mm (stainless steel).
– Gas Temperature : 200 °C max (optionally 600 °C max).
– Oven Temperature : 120 °C
– Calculation Sheet : in Excel® (supplied)
– Packaged weight : 120 kg
– Number of packages : 5
 

 

Key Points

– This device must only be used by an operator trained in a specific methodology. The flow speed at the sampling point must be known and communicated to CleanAir Europe for assessing the feasibility of the project.

Heavy Metal item 311M

It can sample heavy metals.

It is a variant of the base model (dust) and consists of nozzles and a Pyrex or quartz duct with a compartment containing 7 impingers.
 

Specification

– Sampling flow rate : depending on the isokinetic parameters < 40 liters/min
– Length of probe : 0.6 m, 0.9 m, 1.2 m, 1.5 m, 1.8 m, stainless steel (specify length with order).
– Umbilical : 7.5m, 15m or 30 m (specify with order).
– Set of Nozzles : 3, 4, 5, 7, 9 and 12 mm (stainless steel).
– Gas Temperature : 200 °C max (optionally 600 °C max).
– Oven Temperature : 120 °C
– Calculation Sheet : in Excel® (supplied)
– Packaged weight : 120 kg
– Number of packages : 5

 

Key Points

– This device must only be used by an operator trained in a specific methodology. The flow speed at the sampling point must be known and communicated to CleanAir Europe for assessing the feasibility of the project.

PM 10/2,5 item 345 et 346

This cyclone separator is mounted at the end of the sampling probe in the chimney.

The sample is taken at a constant flowrate and particles larger than 10 μm (aerodynamic diameter) are eliminated. Particle mass is determined by weighing after drying.

The PM10/2.5 model samples particles smaller than 10 μm, i.e. those between 10 and 2.5 μm and particles smaller than 2.5μm.

Specification

– Set of 11 nozzles : 3.45 to 9.9 mm
– Spec : PM10 / 2.5
– Filter Diameter : 63 mm (not supplied)
– Packaged weight : 5.8 kg
– Number of packages : 1