ETG MCA 100 BIO-P ITEM 546 AMBIENT AIR AND PROCESS

This system is suitable for the measurement of the biogas in a single point, and performs continuous measurement of CH4, CO2,CO, O2 and H2S.

The MCA 100 P is the portable version of the MCA 100 Bio and has been designed for those who need to monitor the composition of the biogas at several points.

The analyzer is mounted inside a hard case impact resistant, equipped with touch screen monitor ( 5.7”), capable of recording up to 1000 data analysis, has an autonomy of 8 hours.

 

CH4, CO and CO2 board NDIR based

 

O2, H2S board electro chemical cell based

 

Specifications

– Ranges :

  • CH4 0…100 vol% (NDIR)
  • CO2 0…50% (NDIR)
  • CO 0…1% (NDIR)
  • O2 0…25 vol% (ECD)
  • H2S 0…5000 ppm (ECD)
  • Calorific Value

– Unit : %, PPM

– Data logging : Yes

– Software : Yes

– Internal pump : Yes

– Analog output : No

– Weigth : 8 kg

– Number of box : 1

– Recommended for use with: 822

Key Points :

– Real-Time analysis

– Remotable by Ethernet

– Trend of the last 30 minutes for each measured compound

– Dust filter on sample line

– Sampling pump

– Data Download .csv format by USB port

 

 

 

Q-Track item 452

This portable device measures carbon monoxide in ambient air with an electrochemical cell and measures carbon dioxide using non-dispersive infra-red detection.

It can also indicate temperature and relative humidity.

It operates on battery and mains power.

Specification

– Measurement Range : 500 ppm CO, 5000 ppm CO2
– Units : ppm, °C, %
– Temperature : from 0°C to +50°C
– Relative Humidity : 0 – 95%
– Detection Limit : 1 ppm CO, 1 ppm CO2
– Data Storage : 15000 measurement points
– Communication interface : yes
– Software : yes
– Packaged weight : 6 kg
– Number of packages: 1

 

Points clés

– Weight for user : 1 kg
– Powered by 4 AA batteries (1 set in the device, 1 set of new batteries supplied)

GT 521 item 455

This laser diode and isokinetic particle counter will classify particles by size (2 channels) after one minute of sampling (other time periods can be set).

It can be used for checking clean rooms, hospitals, nurseries, effectiveness of filtration systems and ambient air quality.

Specification

– Measurement Range : 0 to 3,106 particles per cubic foot.
– Unit : particles per cubic foot
– Particle classification : 2 channels adjustable between 0.3 μm and 5μm.
– Data storage : yes
– Internal Pump : yes
– Communication interface : yes
– Software : yes
– Packaged weight : 5 kg
– Number of packages : 1

 

Key Points

– Weight for user : 1.1 kg

Gilair 5 item 462

This personal sampling pump is used to sample various pollutants, dust, VOCs and hydrocarbons. It is used in ambient air.

The GilAir-5 sampling pump can be used for constant flows from 0.5 to 5 l/min. A low flow rate module can be added for lower flows, for either constant flow or constant pressure.

The Gilair 5 is used in occupational health and safety, the chemical industry, petrochemical industry, interior air measurements and the pharmaceutical industry.

Specification

– Intrinsic safety
– Adjustable flowrate from 850 cc/min to 5000 cc/min and 20- 500 cc with low flow module.
– Battery Life: 8 hours approx.
– Packaged weight for 6 pumps: 7.5 kg, 1 case.
– Weight of a pump: 0.7 kg with charger
– Recommended for use with: 475

 

 

Aircon II item 466

This high volume sampling system provides an extended range of flowrates.

Designed to collect particles in a factory, it is completely programmable so that you can perform sampling without operator intervention for up to 99 hours.

It can be used for constant flowrates from 2 to 30 liters per minute.

The Aircon II is used in occupational health and safety, the chemical industry, petrochemical industry, interior air measurements and the pharmaceutical industry. It is especially used for sampling asbestos at 7 l/min.
 

Specification

– Completely programmable
– Flowrate adjustable from 2 l/min to 30 l/min.
– Temperate Ranges :
   Operating : -20  C to +45 °C
   Storage : -40 °C to +45 °C
– Instantaneous pressure
– Instantaneous “fault” function
– Battery Life : 8 hours
– Packaged weight : 10 kg

 

C.A.Th.I.A item 470

This device is used for special sampling, particularly asbestos.
The C.A.Th.I.A head is an Alveolar Inhalable Thoracic Aerosol Sensor.
The filter is held between 2 rings which are part of the filter cassette covers so that the whole filter assembly can be removed. The operator does not have to handle the filter at all when installing and removing it.
The machine is supplied with a tripod on request.
 

Specification

– Three selectors : alveolar, inhalable and thoracic
– Flowrate 7l/min
– Filter Cassette-Holder for 37 mm diameter filters
– Meets EN 481, ISO 7708 and FD CEN/TR 15230 standards
– Thoracic selector meets standard NFX 43-050
– Recommended for use with : 466, 475

 

TVA 1000 item 523

This portable device measures VOCs in ambient air.
The TVA 1000 uses flame ionization (cold FID) to measure volatile organic compounds in ambient air.
It can be supplied with a PID built in (optional).

It is particularly suited for identifying leaks.

Specification

– Measurement Range : 0 – 50,000 ppm methane (linear up to 10,000 ppm).
– Units : ppm
– Detection Limit : 0.1 ppm of hexane
– Fuel gas : H2 (refillable cylinders, refill kit supplied)
– Data Storage : 4000 measurement points
– Internal Pump : yes
– Communication interface : yes
– Software : yes
– Packaged weight : 17 kg
– Number of packages : 1
– Recommended for use with : 928, 925MA, 927

 

Key Points

– Approval : EEx ib m IIC T4 (KEMA Certificate No. Ex-94.C.7162)
– Weight for user : 6 kg
– Response factors provided for 77 volatile organic compounds (contact us)
– Battery Life : 8 hours approx
– Life of an H2 cartridge: 4 to 6 hours
 

Gas 3000 item 547

This portable analyzer can measure CH4, CO2, O2 and H2S for biogas applications.

The GAS 3000 uses infra-red detection for CH4 and CO2 and electrochemical cells for O2 and H2S.

It is used to measure gas in the soil, to monitor landfills and waste processing sites and for biogas quality control.

It is supplied with a probe and a hydrophobic filter.

Specification

– Ranges : 0 – 100% CH4, 0 – 50% CO2, 0 – 25% O2, 0 – 5000 ppm H2S
– Units : %, ppm
– Internal Pump : yes
– Data Storage : yes
– Communication interface : yes
– Software : yes
– Recommended for use with : 622

Key Points

– Battery Life : 4 to 6 hours

Jerome J405 item 560

This portable machine will measure mercury vapor in ambient air.

The Jerome J405 uses gold film detection.

It can be used to monitor mercury spillages at hazardous waste sites, dumps, water conduits and so on.
There is very little interference compared to a UV detector.
 

Specification

• Measurement Range : 0.5 – 999 μg/m3
• Unit : μg/m3
• Ambient Air: 0 °C- 45 °C on non-condensable and non-explosive gas.
• Flowrate : 750 cc/min
• Detection Limit : 0.5 μg/m3
• Data Storage: 20000 measurement points
• Internal Pump : yes
• Communication interface : yes
• Software : yes

 

Key Points

• Weight for user : 2.5 kg
• Charger power supply (can operate on a 12V battery, not supplied)
• Automatic zero calibration