Return to the previous page

Servomex 1400 item 113

This analyzer measures the CO2 levels in emissions.
The Servomex 1400 uses the principle of infrared absorption.
It is used after a gas preparation system.
Set-points can be pre-defined and an alarm can be sounded if they are exceeded. A flow-rate alarm is also available.
The 0 – 10 V analog output can be used to record the data on a plotter.

Specification

– Measurement Range : 0 – 20%, 0 – 25%
– Unit : %
– Detection Limit : 0.1%
– Analog output : 0 – 10V
– Data storage : None (a data recorder can be added).
– Internal pump : No (an external pump can be added).
– Packaged weight : 15 kg
– Number of package : 1
– Recommended for use with : 846, 832, 820, 812, 622, 924

Key Points

– 19’’ rack, 4U
– 2 alarms at 2 thresholds

Q-Track item 452

This portable device measures carbon monoxide in ambient air with an electrochemical cell and measures carbon dioxide using non-dispersive infra-red detection.

It can also indicate temperature and relative humidity.

It operates on battery and mains power.

Specification

– Measurement Range : 500 ppm CO, 5000 ppm CO2
– Units : ppm, °C, %
– Temperature : from 0°C to +50°C
– Relative Humidity : 0 – 95%
– Detection Limit : 1 ppm CO, 1 ppm CO2
– Data Storage : 15000 measurement points
– Communication interface : yes
– Software : yes
– Packaged weight : 6 kg
– Number of packages: 1

 

Points clés

– Weight for user : 1 kg
– Powered by 4 AA batteries (1 set in the device, 1 set of new batteries supplied)

VA 3000 item 154

This analyzer measures the N2O levels in emissions.
It uses the principle of infra-red correlation (NDIR) to measure nitrous oxide (N2O) in emissions and in specific processes. Fitted with a CO2 filter to eliminate interference caused by this gas.
It is used after an NO2 converter gas treatment system with a yield of greater than 97%.

Specification

– Measuring range : programmable from 0 – 500 ppm to 0 – 5000 ppm
– Units : ppm, mg/m3
– Detection Limit : 1 ppm
– Internal Pump : yes
– Data storage : none (a data recorder can be added).
– Analog output : yes (0 – 1V or 4 – 20 mA)
– Communication interface : yes
– Packaged weight : 18 kg
– Number of packages : 1
– Recommended for use with : 846, 832, 820, 622, 917.

Key Points

– Reduced heat-up time
– No CO2 interference (optical CO2 filter), reduced CO interference
– Not sensitive to vibration
– Menus for setting parameters and for optical and electronic diagnostics
– Automatic range switchover

Gas 3000 item 547

This portable analyzer can measure CH4, CO2, O2 and H2S for biogas applications.

The GAS 3000 uses infra-red detection for CH4 and CO2 and electrochemical cells for O2 and H2S.

It is used to measure gas in the soil, to monitor landfills and waste processing sites and for biogas quality control.

It is supplied with a probe and a hydrophobic filter.

Specification

– Ranges : 0 – 100% CH4, 0 – 50% CO2, 0 – 25% O2, 0 – 5000 ppm H2S
– Units : %, ppm
– Internal Pump : yes
– Data Storage : yes
– Communication interface : yes
– Software : yes
– Recommended for use with : 622

Key Points

– Battery Life : 4 to 6 hours

AP 360 item 220

This machine measures CO in the open air to determine air quality or to set a zero baseline before building construction work is carried out.
The Horiba AP360 uses NDIR detection.

Specification

– Measurement Range : 0 – 100 ppm
– Units : ppm
– Detection Limit : 0.02 ppm
– Data storage : Yes
– Internal pump : yes
– Analog output : Yes
– Communication interface : yes
– Packaged weight : 20 kg
– Number of packages : 1
– Recommended for use with : 622, 923, 927.

Key PointsPoints clés

– 19’’ rack

Ultramat 23 item 120 121 122

This continuous gas analyzer can measure the CO and CO2 in emissions simultaneously.
The Ultramat 23 uses the principle of infrared absorption.
Several versions with different ranges are available.
The machine can be delivered with its calibration certificate on request (depending on the range).

Specification

– Measurement Ranges:
Item 120 :  0 – 0.05% and 0 – 2.5% CO and 0 – 5% and 0 – 25% CO2
Item 121 : 0 to 1000 ppm CO and 0 to 10,000 ppm CO2 (full scale)
Item 122 : 0 – 10% and 0 – 50% CO and 0 – 1% CO2
– Units : %, mg/m3
– Detection limit : 50 ppm CO / 500 ppm CO2
Item 120 : 50 ppm CO / 500 ppm CO2
Item 121 : 1 ppm CO / 1 ppm CO2
Item 122 : 0.05% ppm CO / 500 ppm CO2
– Data storage : none (a data recorder can be added).
– Internal Pump : yes
– Analog Output : 0 – 10 V
– Packaged weight : 15 kg
– Number of packages : 1
– Recommended for use with : 846, 832, 820, 622, 923, 924.

Key Points

– 19’’ rack, 4U

48C item 124

This analyzer measures CO levels in emissions and in ambient air.
The Thermo 48C uses the principle of infra-red correlation. This technique uses an interference filter to eliminate the effect of the other gases present. It is used on gas which is dry, clean and at room temperature.
It is ideal for very low concentrations.

Specification

– Measurement range : programmable from 0 – 1 ppm to 0 – 10,000 ppm.
– Units : ppm, mg/m3
– Detection Limit : 0.04 ppm
– Data storage : 5000 measurement points
– Internal Pump : yes
– Analog Output : 0 – 10 V
– Communication interface : yes
– Software : yes
– Packaged weight : 28 kg
– Number of packages : 1
– Recommended for use with : 846, 832, 820622, 923

Key Points

– Menus for setting parameters and optical and electronic diagnostics
– Automatic range switchover
– Fast response time
– Linearity specific to scales
– Excellent linearity and reduced sensitivity to interference